/!\ Risk of bottle explosion.
/!\ First disconnect the mains plug.
This symptom is generally caused by a problem to the pressure regulator (fault or low temperature) or by a crack in the water bottle.
- Stop using your device immediately and disconnect it from the grid.
- Check the bottle in use for crack or any damage and, if the case, replace it with a new one.
- Check the supplied air pressure: it shall be 4.5 bar minimum. 6 bar ideally.
- If the device temperature is below 10°C (device too cold), wait for it to warm-up at ambient temperature and then reconnect to the power grid and switch it on again.
- If the device temperature is over 10°C, or the problem recurs, definitively stop using it and contact EMS aftersales service.